1. What Is Strain Rate Control?

Strain rate control is a testing method in which the deformation rate of a specimen is controlled or estimated during a mechanical test. It is particularly important in tensile testing because many materials exhibit changes in mechanical behavior when the rate of deformation changes.

A Universal Testing Machine equipped with suitable control software and strain measurement equipment can regulate the loading process according to a specified strain rate. Instead of controlling only the movement of the crosshead, the system can use specimen deformation information to adjust machine movement.

This capability is important in applications where tensile strength, yield strength, proof strength, elongation, and other mechanical properties are sensitive to the rate of deformation.

1.1 What Is Strain Rate?

Strain rate describes how quickly strain changes with time. It is commonly expressed in reciprocal seconds (s⁻¹) or another unit appropriate to the test method.

Strain itself represents relative deformation compared with an original reference length. Strain rate therefore describes the speed at which this relative deformation develops during loading.

1.2 Why Strain Rate Is Different from Machine Speed

The crosshead speed describes the movement of the testing machine's moving crosshead. Strain rate describes deformation of the specimen over a defined measurement length.

These values are related, but they are not always identical. Machine compliance, grip deformation, specimen geometry, alignment, and other mechanical effects can cause the actual specimen deformation rate to differ from the crosshead movement rate.

1.3 Where Strain Rate Control Is Used

Strain rate control is commonly associated with metallic tensile testing and standards that define testing rates based on specimen deformation. It can also be relevant in advanced research on polymers, composites, and other materials whose properties depend on deformation rate.

2. Strain Rate vs. Crosshead Speed

2.1 Understanding Crosshead Speed

Crosshead speed is the linear movement speed of the Universal Testing Machine's moving crosshead. It is usually expressed in millimeters per minute (mm/min) or another unit of displacement per time.

For a simple tensile test, increasing the distance between the grips causes the specimen to extend. However, not all crosshead movement is necessarily converted into deformation within the specimen's gauge section.

2.2 Understanding Specimen Strain

Specimen strain is calculated from the change in a defined gauge length relative to its original length. An extensometer or another suitable measurement device can measure the deformation used to calculate strain.

Because strain is based on specimen deformation rather than total machine movement, it can provide a more direct indication of how quickly the material is being stretched.

2.3 Machine Compliance

Every mechanical testing system has some degree of compliance. The load frame, grips, fixtures, connections, and other components may deform slightly under force.

When the crosshead moves, part of that movement can be associated with deformation of the testing system rather than deformation of the specimen gauge section.

2.4 Why Two Machines Can Produce Different Results

Two Universal Testing Machines may be programmed to use the same crosshead speed but have different structural stiffness, grip arrangements, and mechanical compliance.

As a result, the specimen deformation rate may not be identical on both machines. This difference can influence measured mechanical properties, particularly when the material is sensitive to strain rate.

2.5 When Crosshead Speed Is Sufficient

Crosshead-speed control may be suitable when the applicable test procedure permits it and when the required specimen deformation rate can be estimated reliably.

For procedures requiring direct specimen strain-rate control, the testing machine should use an appropriate strain measurement and feedback arrangement.

3. Why Strain Rate Control Matters

3.1 Material Behavior Depends on Deformation Rate

Some materials respond differently when they are deformed slowly or rapidly. Their measured strength, stiffness, yield behavior, and elongation may change with the loading rate.

This behavior is commonly described as strain-rate sensitivity. The magnitude and direction of the effect depend on material composition, microstructure, temperature, and the relevant deformation mechanism.

3.2 Accurate Yield and Proof Strength Measurement

Yield strength and proof strength are important parameters in metallic material testing. Their determination can be influenced by the strain rate used during the relevant part of the tensile test.

Controlling the strain rate according to the applicable standard helps reduce variation caused by differences in testing conditions.

3.3 Improving Interlaboratory Comparability

When different laboratories use different control methods or test rates, their results may not be directly comparable.

Using the same applicable standard, control method, specimen geometry, measurement range, and test conditions helps improve the consistency of comparisons.

3.4 Reducing Measurement Uncertainty

Strain-rate control can reduce the influence of some machine-compliance effects when the control system uses suitable feedback from the specimen.

It does not eliminate every source of uncertainty. Force measurement, extensometer verification, specimen dimensions, alignment, and other factors remain important.

3.5 Supporting Advanced Material Research

Researchers may need to investigate the relationship between deformation rate and mechanical response. A programmable Universal Testing Machine can support tests at different strain-rate conditions when the required control and measurement capabilities are available.

4. Closed-Loop and Open-Loop Control

4.1 What Is Closed-Loop Control?

Closed-loop control uses feedback from a measurement device to adjust the machine's movement. In strain-rate control, the feedback may come from an extensometer measuring specimen deformation.

The controller compares the measured strain rate with the target value. It then adjusts the crosshead movement to reduce the difference between the measured and target rates.

4.2 How Closed-Loop Strain Control Works

At the beginning of the controlled stage, the system measures specimen strain over time. The controller calculates the current deformation rate and compares it with the programmed target.

If the measured strain rate is below the target, the controller can increase crosshead movement. If the measured rate is above the target, it can reduce movement according to the control algorithm and machine limits.

4.3 What Is Open-Loop Control?

Open-loop control does not continuously use direct feedback from the specimen strain measurement to regulate the target strain rate. Instead, the machine uses a calculated or estimated movement rate based on specimen dimensions and the specified deformation rate.

For certain standardized procedures, an estimated strain rate based on crosshead separation and specimen geometry may be permitted.

4.4 Differences Between the Two Methods

Closed-loop control responds to measured specimen deformation, while open-loop control relies on a predetermined movement calculation or estimate.

Closed-loop control can compensate for certain changes in machine compliance more directly, but it requires reliable feedback, suitable control settings, and appropriate extensometer performance.

4.5 Control Stability

Closed-loop control requires a properly configured control system. Excessive controller gain, unstable feedback, poor extensometer attachment, or inappropriate safety settings can lead to unwanted machine movement.

Control parameters should therefore be configured by qualified personnel following the equipment manufacturer's procedures.

5. The Role of Extensometers

5.1 What Is an Extensometer?

An extensometer is a device used to measure deformation over a defined gauge length of a specimen. It provides strain information that can be used for material-property calculations and, in suitable systems, closed-loop strain-rate control.

5.2 Contact Extensometers

Contact extensometers attach to or contact the specimen at defined measurement points. They measure the change in distance between those points during loading.

They are widely used in tensile testing when the specimen geometry and deformation range are suitable for the instrument.

5.3 Automatic Extensometers

Automatic extensometers can attach to and detach from the specimen through a controlled mechanism. Depending on the model and configuration, they can reduce manual handling and improve the efficiency of repeated tensile tests.

Some systems can also be integrated with the Universal Testing Machine controller to provide strain feedback.

5.4 Non-Contact Extensometers

Video and optical extensometers measure specimen deformation without physical contact. They can be useful when contact measurement may damage the specimen, interfere with deformation, or become impractical at large elongations.

Non-contact systems still require appropriate calibration, image quality, measurement range, and compatibility with the machine's control system if they are intended for closed-loop operation.

5.5 Extensometer Gauge Length

The gauge length is the defined initial measurement distance on the specimen. The selected gauge length should match the test method and specimen geometry.

Incorrect gauge length configuration can affect calculated strain and strain-rate values.

5.6 Extensometer Verification

The extensometer should be suitable for the strain range in which the required mechanical properties are determined. Verification status, resolution, travel, attachment method, and measurement performance should be checked before testing.

6. Strain Rate Control in ISO 6892-1

6.1 Scope of ISO 6892-1

ISO 6892-1 specifies the method for tensile testing of metallic materials at room temperature. It includes provisions for testing rates and defines approaches based on strain rate and stress rate.

For laboratories testing metals, understanding the applicable control method is important because the selected rate can affect the measured mechanical properties.

6.2 Method A: Strain Rate-Based Testing

Method A is based on controlling or estimating the strain rate over the relevant specimen length. The purpose is to reduce variation in the testing rate when strain-rate-sensitive properties are determined.

The method includes different approaches, including direct closed-loop strain-rate control and an estimated strain-rate approach based on crosshead movement and specimen geometry.

6.3 Method A1: Closed-Loop Strain Rate Control

In Method A1, the strain rate is controlled using feedback from an extensometer measuring specimen deformation.

The Universal Testing Machine adjusts crosshead movement based on the measured strain response. This approach can reduce the influence of testing-system compliance during the relevant measurement stage.

6.4 Method A2: Estimated Strain Rate

Method A2 uses an estimated strain rate based on crosshead separation rate and specimen geometry. It is an alternative approach when direct closed-loop strain-rate control is not available or is not used.

The calculation and operating conditions must follow the applicable edition of ISO 6892-1.

6.5 Method B: Stress Rate-Based Testing

Method B is based on controlling the rate of increase of stress during the applicable stage of the test. It is distinct from strain-rate control and should not be treated as the same control mode.

The laboratory should select the method required by the standard, product specification, or customer requirement.

6.6 Rate Changes During the Test

Some procedures involve different testing rates at different stages of the tensile test. The transition between rates should be managed so that the change does not introduce unwanted discontinuities into the stress-strain curve.

7. Strain Rate Control in ASTM E8/E8M

7.1 ASTM E8/E8M Overview

ASTM E8/E8M covers tension testing of metallic materials at room temperature. It includes requirements and options for controlling the testing rate during the determination of tensile properties.

Laboratories should refer to the current edition of ASTM E8/E8M for the applicable control methods and their specific conditions.

7.2 Control Method Selection

ASTM E8/E8M describes different approaches for controlling test speed or deformation rate during specified stages of the test. The selected approach can influence the measured result, particularly for materials that are sensitive to deformation rate.

7.3 Extensometer Feedback

Where closed-loop strain control is required, the extensometer signal provides feedback to the testing machine. The controller uses that information to maintain the target deformation rate within the applicable operating requirements.

7.4 Crosshead Speed-Based Control

Crosshead speed can be used in the situations permitted by the standard. The speed is determined using the relevant specimen dimensions and specified rate.

However, crosshead speed and actual specimen strain rate should not be assumed to be identical in every mechanical configuration.

7.5 Consistency Between Laboratories

When comparing results, laboratories should record the control method and test conditions. Using different methods can produce different results, particularly when the tested material exhibits significant strain-rate sensitivity.

8. Universal Testing Machine System Configuration

8.1 Stiff Load Frame

A suitable load frame should provide adequate stiffness and stable movement for the intended tensile test. Structural compliance can affect the relationship between crosshead movement and specimen deformation.

8.2 Servo Drive System

A controlled servo drive can provide accurate crosshead movement and support programmed changes in test speed. The drive system should be suitable for the required loading range and control mode.

8.3 Force Measurement System

The load cell measures tensile force and provides the force information needed for stress calculation. It should have suitable capacity and verified measurement performance over the expected test range.

8.4 Extensometer and Feedback Interface

For closed-loop strain control, the extensometer must be compatible with the testing machine controller. The measurement signal should be stable and suitable for real-time control.

8.5 Control Software

The software should allow the operator to configure the required test method, strain rate, control transitions, specimen dimensions, and data recording settings.

8.6 Safety Limits

Safety limits are important when using closed-loop control. The system should include appropriate travel limits, force limits, extensometer range protection, and emergency stop functions.

9. Strain Rate Control During a Tensile Test

9.1 Specimen Preparation

Prepare the specimen according to the applicable test standard. Measure the required dimensions and confirm that the specimen is free from defects that could affect the tensile result.

9.2 Install the Specimen

Mount the specimen in suitable grips and align it with the machine loading axis. Incorrect alignment can introduce bending and affect the measured strain.

9.3 Attach the Extensometer

Install the extensometer at the required gauge length. Confirm that the device is securely attached and has sufficient measurement range for the intended test stage.

9.4 Configure the Control Mode

Select the control mode required by the applicable standard. Enter the specified strain rate or estimated rate, specimen dimensions, and relevant transition conditions.

9.5 Check Feedback and Zero Values

Confirm that the extensometer signal is stable and that the force and strain channels are correctly configured. The measurement system should be checked before applying significant tensile load.

9.6 Start the Test

Start the tensile test and monitor the machine's response. During closed-loop operation, the controller uses the strain feedback signal to adjust crosshead movement.

9.7 Change Test Rate When Required

If the method specifies different rates for different stages, the controller should perform the transition according to the configured procedure.

9.8 Complete the Test and Review Results

After the test, review force, strain, displacement, and the stress-strain curve. Check whether the rate control operated correctly throughout the relevant stages.

10. Factors Affecting Strain Rate Accuracy

10.1 Machine Stiffness

Machine stiffness affects how much crosshead movement is absorbed by deformation of the testing system. A more compliant system may create a larger difference between crosshead movement and specimen deformation.

10.2 Grip Compliance

Grips and fixtures can deform under load. Their mechanical behavior can influence the actual deformation experienced by the specimen.

10.3 Specimen Alignment

Misalignment can introduce bending and non-uniform deformation. Correct alignment is important for reliable strain measurement and control.

10.4 Extensometer Performance

Extensometer resolution, accuracy, gauge length, attachment stability, and measurement range all influence the quality of strain feedback.

10.5 Controller Settings

Incorrect controller parameters can cause unstable movement, delayed response, or overshoot. Control settings should be established according to the manufacturer's recommendations and validated for the testing configuration.

10.6 Material Behavior

Discontinuous yielding, serrated yielding, necking, and localized deformation can complicate strain-rate control. The measurement method should be appropriate for the material response and the relevant stage of the test.

10.7 Data Acquisition Frequency

The acquisition system should record strain and force data frequently enough to characterize the test response. Insufficient sampling can make it difficult to identify rapid changes or short-duration events.

11. Common Problems and Troubleshooting

11.1 Strain Rate Does Not Reach the Target

Possible causes include insufficient drive performance, unsuitable controller settings, extensometer signal problems, excessive system compliance, or an incorrectly configured test method.

Check the control configuration, feedback signal, specimen setup, and machine operating limits before repeating the test.

11.2 Unstable Crosshead Movement

Unstable movement during closed-loop control may be caused by poor feedback quality, inappropriate control parameters, extensometer slippage, or mechanical interference.

Review the controller settings and verify the extensometer installation before continuing.

11.3 Extensometer Slippage

If an extensometer moves relative to the specimen, the measured strain may no longer represent actual gauge-length deformation.

Check the attachment method and ensure that the extensometer is suitable for the specimen surface and geometry.

11.4 Incorrect Gauge Length

An incorrect gauge length can produce inaccurate strain and strain-rate calculations. Confirm that the entered value matches the actual measurement configuration.

11.5 Differences Between Test Machines

Different load-frame stiffness, grips, and control configurations can lead to differences in specimen strain rate when the machines are operated at the same crosshead speed.

For meaningful comparison, laboratories should use consistent control methods and verify the actual measurement conditions.

11.6 Sudden Movement During Control Transition

Unexpected crosshead movement may occur if the transition between control modes is not configured properly or if the feedback signal becomes unstable.

Use suitable transition settings and safety limits, and follow the machine manufacturer's operating instructions.

12. How to Select a Universal Testing Machine with Strain Rate Control

12.1 Identify the Applicable Test Standards

Determine whether the laboratory needs to perform ISO 6892-1, ASTM E8/E8M, or other material testing procedures. The standard determines the required control method and measurement conditions.

12.2 Check Closed-Loop Control Capability

If direct strain-rate control is required, confirm that the testing machine controller supports closed-loop control using an appropriate extensometer signal.

12.3 Select a Suitable Extensometer

The extensometer should match the specimen geometry, gauge length, strain range, and required accuracy. Confirm whether automatic attachment or non-contact measurement is needed.

12.4 Evaluate Frame Stiffness

Machine stiffness is relevant when controlling specimen deformation rate. The frame, grips, and fixtures should provide an appropriate mechanical configuration for the intended application.

12.5 Check Servo Drive Performance

The drive system should support stable movement and the required speed range. It should also respond appropriately to the controller's feedback commands.

12.6 Review Software Functions

Confirm that the software supports strain-rate programming, control-mode selection, test-stage transitions, data recording, stress-strain curves, and report generation.

12.7 Confirm Calibration and Verification

Force measurement and extensometer measurement should be calibrated or verified as required. The laboratory should maintain documentation for the relevant measurement channels.

12.8 Consider Future Testing Requirements

For laboratories testing multiple metal grades or developing new materials, flexible control software and interchangeable measurement accessories can support future test requirements.

13. Frequently Asked Questions

13.1 What Is Strain Rate Control in a Universal Testing Machine?

Strain rate control is a method of regulating the rate at which a specimen deforms during testing. It can use extensometer feedback or an estimated rate based on machine movement and specimen geometry, depending on the selected procedure.

13.2 What Is the Difference Between Strain Rate and Crosshead Speed?

Crosshead speed describes machine movement, while strain rate describes the rate of deformation over a defined specimen length. The two are related but may differ because of machine and fixture compliance.

13.3 What Is Closed-Loop Strain Control?

Closed-loop strain control uses feedback from a strain measurement device, commonly an extensometer, to adjust the movement of the Universal Testing Machine and maintain the target strain rate.

13.4 Does Strain Rate Control Require an Extensometer?

Direct closed-loop strain-rate control generally requires a suitable strain measurement feedback device. Some standards also permit an estimated strain-rate approach based on crosshead movement and specimen geometry.

13.5 Why Does Strain Rate Affect Tensile Test Results?

Some materials are sensitive to deformation rate. Their yield strength, tensile strength, elongation, or other mechanical properties may change when the strain rate changes.

13.6 Can the Same Machine Perform Both Strain Rate and Crosshead Speed Control?

Many modern Universal Testing Machines support different control modes, provided that the controller, software, and measurement accessories are configured for the required method.

13.7 Is Strain Rate Control Important for Metal Testing?

Yes. It is particularly important for metallic tensile testing procedures that define testing rates based on specimen strain or estimated strain rate.

14. Conclusion

Strain rate control is an important capability for Universal Testing Machines used in advanced tensile testing. It allows laboratories to regulate or estimate specimen deformation rate rather than relying exclusively on crosshead movement.

The difference between strain rate and crosshead speed becomes especially important when machine compliance, grip deformation, and material rate sensitivity influence the test response. Closed-loop control using an extensometer can provide direct specimen feedback, while permitted open-loop methods can estimate the deformation rate from machine movement and specimen geometry.

For reliable results, laboratories should select a suitable testing machine, extensometer, load cell, control system, and software configuration. The selected method must also comply with the applicable testing standard and the laboratory's measurement requirements.

A properly configured Universal Testing Machine with strain-rate control can support more consistent metallic tensile testing, improve the quality of mechanical-property data, and provide valuable capabilities for material research and industrial quality control.

Is this blog helpful to you?
0
0
0 Comments
Leave a Comment
Your email address will not be published. Required fields are marked *
😍
😜
😳
😌
😄
😘
😝
😒
😃
😚
😚
😛
😟
😧
😀
😉
😓
😱
😤
😣
😂
😥
😩
😠
😢
😭
😰
😨
😡
😆
😪
😅
😐
😇
😋
😴
👿
😕
😏
😷
😵
😟
😮
😯
😑
👧
👴
😧
😬
😾
👶
👱
👵
👸
🙀
👺
👦
👩
👨
😽
😿
🙈
💩
💥
💤
😼
😹
🙉
🔥
✨
💦
👎
✌
👆
👈
💪
💹
👍
👊
💴
💶
💷
💸
👉
💵
🙏
🌎
🏧
👏
💳
👇
💑
🙆
🙅
💁
👫
👭
🙎
🙇
👑
👔
Submit Comment
Contact Us Now
The Best Testing Machine Manufacturer & Supplier in China - WTS
East Road, 200 meters south of Jixi Used Car Market, Jingshi West Road, Shizhong District, Jinan City, Shandong Province, China
You can trust us
As a professional China testing machine manufacturer, we are dedicated to continuous innovation, ensuring our customers receive superior products and services.
Contact Us
© 2025 Jinan Wenteng Testing Instrument Co., Ltd.        SiteMap.html    SiteMap.xml    Terms of Service      Privacy Policy
Marketing Support by Globalsir
Enter your inquiry details, We will reply you in 24 hours.
Name can't be empty
E-mail can't be empty
Company can't be empty
Phone can't be empty
Products can't be empty
Message can't be empty
Verification code error
code
Clear